Draeger Tube, Monostyrene 10/A 10-200ppm
Product ID: 6723301
Price: $86.95
Availability: Email for Availability (info@nafeco.com)
 
Order Qty:
   
 
 
 

Draeger Tube - Styrene 10/a
Order No. 67 23 301

Application Range
Standard Measuring Range 10 to 200 ppm
Number of Strokes n: max. 15
Time for Measurement: max. 3 minutes
Standard Deviation: ± 15 to 20%
Color Change: white to pale yellow
   
Ambient Operating Conditions
Temperature: 15 to 40° C ~ 59 to 104° F
Absolute Humidity: < 15 mg H2O / L
   
Reaction Principle
C6H5-CH=CH2 + H2SO4 → pale yellow reaction product
 
Cross Sensitivity
Other organic compounds which tend toward polymerization (e. g. butadiene) are indicated, but with different sensitivities. It is impossible to measure monostyrene in the presence of these compounds.

View the entire Draeger Tubes & CMS Handbookk