Draeger Tube, Sulfur Dioxde 1/A 1-25ppm

$86.00
Availability: Call for Availability
Product Code: CH31701
Brand: DRAEGER SAFETY INC.
+

Draeger Tube - Sulfur Dioxide 1/a
Order No. CH 31 701

Application Range
Standard Measuring Range 1 to 25 ppm
Number of Strokes n: 10
Time for Measurement: approx. 3 minutes
Standard Deviation: ± 10 to 15%
Color Change: grey blue to white
   
Ambient Operating Conditions
Temperature: 15 to 25° C ~ 59 to 77° F
Absolute Humidity: 3 to 20 mg H2O /L
   
Reaction Principle
SO2 + I2 + 2 H2O → H2SO4 + 2 HI
 
Cross Sensitivity
Hydrogen sulfide in the TLV range is retained in the pre-layer and does not interfere. Nitrogen dioxide will shorten the reading.

View the entire Draeger Tubes & CMS Handbookk