Draeger Tube - Toluene 5/b
Order No. 81 01 661

Application Range
Standard Measuring Range 50 to 300 ppm / 5 to 80 ppm
Number of Strokes n: 2 / 10
Time for Measurement: Approx. 2 minutes / Approx. 10 minutes
Standard Deviation: ± 10 to 15%
Color Change: white to pale brown
   
Ambient Operating Conditions
Temperature: 2 to 40° C ~ 35.6 to 104° F
Absolute Humidity: max. 20 mg H2O /L
   
Reaction Principle
Toluene + I2O5 + H2SO4 → I2
 
Cross Sensitivity
10 ppm phenol, 1,000 ppm acetone, 1,000 ppm ethanol and 300 ppm octane are not indicated, xylene (all isomers) and benzene are indicated with the same sensitivity. The discoloration in the presence of p-xylene is violet, and yellowish-green with benzene.