Register to Win the Flying Cross & Vertx Giveaway Package

Draeger Tube, Box of 8 Halogenated Hydrocabons

Draeger Tube, Box of 8 Halogenated Hydrocabons

$340.00
Availability: May direct ship from manufacturer.
Product Code: 8101601

Draeger Tube - Halogenated Hydrocarbons 100/a
Order No. 81 01 601

Application Range
Standard Measuring Range 200 to 2,600 ppm R 113/R 114
100 to 1,400 ppm R 11
200 to 2,800 ppm R 22
1,000 to 4,000 ppm R 134a (on request)
The indication is in mm and must be compared to the calibration data sheets
Number of Strokes n: 3
Time for Measurement: approx. 1 minutes
Standard Deviation: ± 30%
Color Change: blue to yellow / greyish green
Ambient Operating Conditions
Temperature: 0 to 40° C ~ 32 to 104° F
Absolute Humidity: 1 to 15 mg H2O / L
Reaction Principle
a) R113 [pyrolysis]  HCI
b) HCI + pH indicator  yellow reaction product
Cross Sensitivity
Other halogenated hydrocarbons, free halogens and hydrogens halides are also indicated, however, with differing sensitivity. Perchloroethylene is indicated with the same sensitivity as R113.
Additional Information
The tubes become very warm during the measurement. Therefore, these tubes should not be used in potentially combustible atmospheres. A combustible gas monitor should be used to qualify any questionable area before conducting a measurement with one of these tubes.
Share: