Draeger Tube, Hydrogen Fluoride
Product ID: CH30301
Price: $87.95
Availability: Email for Availability (info@nafeco.com)
 
Order Qty:
   
 
 
 

Draeger Tube - Hydrogen Fluoride 1.5/b
Order No. CH 30 301

Application Range
Standard Measuring Range 1.5 to 15 ppm
Number of Strokes n: 20
Time for Measurement: approx. 2 minutes
Standard Deviation: ± 15 to 20%
Color Change: pale blue to pale pink
   
Ambient Operating Conditions
Temperature: 15 to 30° C ~ 59 to 86° F
Absolute Humidity: max. 9 mg H2O / L
   
Reaction Principle
HF + Zr(OH)4/Quinalizarin  [ZrF6]2- + Quinalizarin
 
Cross Sensitivity
In the presence of higher humidity (> 9 mg H2O / L), hydrogen fluoride mist is generated, which cannot be quantitatively indicated by the detector tube (i.e. the indication is too low). Other halogenated hydrocarbons in the TLV range do not interfere.