Draeger Tube, Cyanogen Chloride
Product ID: CH19801
Price: $100.95
Availability: Email for Availability (info@nafeco.com)
 
Order Qty:
   
 
 
 

Draeger Tube - Cyanogen Chloride 0.25/a
Order No. CH 19 801

Application Range
Standard Measuring Range 0.25 to 5 ppm
Number of Strokes n: 20 to 1
Time for Measurement: max. 5 minutes
Standard Deviation: ± 30%
Color Change: white to pink
   
Ambient Operating Conditions
Temperature: 5 to 40° C ~ 41 to 104° F
Absolute Humidity: < 50 mg H2O / L
   
Reaction Principle
a) CICN + pyridine  glutaconaldehyde cyanamide
b) Glutaconaldehyde + barbituric acid  pink reaction product
 
Cross Sensitivity
Cyanogen bromide is indicated as well, but with a different sensitivity. Calibration data is not available.
 
Additional Information
Before performing the measurement the reagent ampoule must be broken and the liquid transferred onto the indicating layer so that it is completely saturated.

View the entire Draeger Tubes & CMS Handbook