Draeger Tube Nickel Tetracarbonyl
Product ID: CH19501
Price: $137.95
Availability: Email for Availability (info@nafeco.com)
 
Order Qty:
   
 
 
 

Draeger Tube - Nickel Tetracarbonyl 0.1/a
Order No. CH 19 501

Application Range
Standard Measuring Range 0.1 to 1 ppm
Discoloration compared to color standard
Number of Strokes n: 20
Time for Measurement: approx. 5 minutes
Standard Deviation: ± 50%
Color Change: yellow to pink
   
Ambient Operating Conditions
Temperature: 0 to 30° C ~ 32 to 86° F
Absolute Humidity: < 30 mg H2O / L
   
Reaction Principle
a) Ni(CO)4 + I2  Nil2 + 4 CO
b) Nil2 + Dimethylglyoxime  pink colored complex
 
Cross Sensitivity
Iron pentacarbonyl is also indicated by a brownish discoloration, however, with less sensitivity. Measurement of nickel tetra carbonyl is not possible in the presence of hydrogen sulfide of sulfur dioxide, since the reading is suppressed. Such a disturbance can be recognised by decoloration of the indicating layer even before the reagent ampoule is opened.
 
Additional Information
After performing the required 20 pump strokes the reagent ampoule must be broken and the liquid carefully drawn onto the indicating layer using the pump.

View the entire Draeger Tubes & CMS Handbookk